No renderer 'timeline' found for mode 'timeline'

Machine Definitions

These are the currently proposed definitions of IHE machines for testing. The definitions here have been provided by Varian and are in xml format.

Machine Name IHE_V120 IHE_V120FFF IHE_ELVHD
Machine Model Varian 23EX
or equivalent
Varian 23EX
or equivalent
Elekta VersaHD
Gantry Angle [0°, 360°) 1) [0°, 360°) 2) [0°, 360°)
Collimator Angle [195°, 165°] [195°, 165°] [0°, 360°)
Dose Rate 100-1000 MU/min 3) 100-1400 MU/min 4) 0-400 MU/min
Electron 9 MeV 9 MeV 9MeV
Photon 6, 18 MV 6, 18 MV 6 MV
MLC
MLC Model Millennium 120 Millennium 120 Agility 160
Pairs 60 60 80
Y Offset -195 to 195 mm -195 to 195 mm -197.5 to 197.5 mm
Width 1-10: 10 mm
11-50: 5 mm
51-60: 10 mm
1-10: 10 mm
11-50: 5 mm
51-60: 10 mm
5 mm
X Jaws
Width (none)
Position (none)
Y Jaws
Width
Position
Couch
Vertical Limits -710 to 460 mm -710 to 460 mm -750 to 500 mm
Longitudinal Limits 150 to 1600 mm 150 to 1600 mm 0 to 800 mm
Lateral Limits -250 to 250 mm -250 to 250 mm -250 to 250 mm
Angle 265-95° 265-95° 260 - 100°
Accessories
Applicators Yes Yes Yes
Compensators Yes Yes Yes
Wedges Yes Yes Yes
Blocks Yes Yes Yes
Compensators Yes Yes ?
XML document V120 XML V120FFF XML ELVHD XML

Wedges

Machine Wedge ID
(300A,00D4)
Wedge Angle
(300A,00D5)
Wedge Orientation
(300A,00D6)
Source-to-Wedge Tray Distance
(300A,00DA)
Orientation
(Varian)
Maximum Field Size
IHE_V120 W60R15U 60º 90º 57.4 cm Right 15 cm x 40 cm
W60L15U 270º Left
W60IN15U In
W60OUT15U 180º Out
W30R30U 30º 90º Right 30 cm x 40 cm
W30L30U 270º Left
W30IN30U In
W30OUT30U 180º Out
EDW20IN 20º N/A In 40 cm x 30 cm
EDW20OUT 180º Out
EDW45IN 45º In
EDW45OUT 180º Out
IHE_S160 3HBW15 FIXME
4HBW45 FIXME
1RW15 FIXME
2VW45 FIXME
IHE_ELVHD MW 60º N/A In 21 cm x 16 cm

Blocks

Machine Block Material ID
(300A,00E1)
Block Tray ID
(300A,00F5)
Source-to-Block Tray Distance
(300A,00F6)
Transmission
(300A,0102)
IHE_V120 Block_IHE User X tray 654 mm 0.01
User E tray 950 mm 0.0

Compensators

Machine Compensator Material ID Compensator Tray ID Source-to-Compensator Tray Distance
(300A,00E6)
IHE_V120 Comp_IHE Usr 1X -4 698 mm

Applicators

Machine Applicator ID
(300A,0108)
Applicator Type
(300A,0109)
Applicator Aperture Shape
(300A,0432)
Applicator Opening
(300A,0433)
Source to Applicator Mounting Position Distance
(300A,0436)
IHE_V120 A10 ELECTRON_SQUARE SYM_SQUARE 100 mm 654 mm
20mm Cone PHOTON_CIRC SYM_CIRCULAR 20 mm 654 mm
1) , 2)
can rotate clockwise from -185° to +185°
3)
Dose rates 800-1000 MU/min indicate SRS primary fluence mode. Do not use for non-SRS beams!
4)
Dose rates 800-1400 MU/min indicate SRS or FFF primary fluence mode. Do not use for non-SRS beams!