IHE Radiation Oncology - MMRO-II Profile Testing

The MMRO-II profile defines the information required for image registration, contouring and display using multiple image series, and the display of the resultant dose calculation on the different image modalities. It requires that these systems support CT, PET, and MRI imaging as a minimum standard of performance. The MMRO-II profile updated the MMRO profile to correct it for changes that occurred in the DICOM standard. For details on the profile, visit the MMRO-II-page.

General Instructions - All Actors

  • The MMRO-II Profile is tested by transferring datasets through a sequence of Actors. Each producer Actor in this sequence is to be tested with three downstream consumers, and each consumer Actor is to be tested using data from three upstream producers.
  • Several datasets are used to test various aspects of image registsration, contouring and display:
    • Datasets 2, 3, and 4 demonstrate handling of non-HFS patient scan positions.
    • Datasets 5, 6, and 10 demonstrate handling of multiple imaging modalities (CT, MR, PET).
    • Datasets 5 and 6 demonstrate handling of hybrid (single frame of reference) and non-hybrid (multiple frame of reference) datasets.
  • Testing guidelines
    1. Each Actor must demonstrate at least one patient with mixed modalities (patient #10) with each test partner.
    2. Each Actor must demonstrate successful exchange of datasets 2, 3, 4 with at least one upstream and one downstream test partner. Be prepared to demonstrate any of these datasets with each peer, (at the judge's request).
    3. Each Actor must demonstrate correct handling of hybrid (shared FoR) and non-hybrid (separate FoR) datasets.
Patient datasets used
# Patient ID Description
2 MMRO15A02vv CT-CT Phantom Scans (HFP)
3 MMRO15A03vv CT-CT Phantom Scans (FFS)
4 MMRO15A04vv CT-CT Phantom Scans (FFP)
5 MMRO15A05vv CT/MR/PET (HFS) Hybrid CT-PET (same FoR), MR in separate FoR
6 MMRO15A06vv CT/MR/PET (HFS) Non-hybrid, CT, PET, MR in separate FoR
10 MMRO15A10vv CT,CBCT/PET (same FoR),MRx2 (HFS)(sagittal)

vv = Vendor code for the Registrator Actor

Note: Dataset #10 includes a Sagittal MR series. Import and display of Sagittal image series is not required to pass testing of the MMRO-II Profile.

Registrator Actor

Preparation

  1. Import images from patient 2,3,4,5,6,10

Testing For each imported patient, the Tester must do the following:

  1. Register all image Series. Remember that the CT Image Series defines the Registered (“Primary”) Frame of Reference.
  2. Export the DICOM Spatial Registration to the Archive.
  3. Create a second rigid registration with a known (noted) offset and export this Spatial Registration to the Archive.
  4. Optional: If possible, also create one contour on the primary image set to aid the Registered Display Actor, and save the RT Structure Set to the Archive.

Then for each peer (Registered Display, Registered Contourer):

  1. Let Registered Display or Registered Contourer import images and registrations from the Registrator
  2. Compare Registrator and Registered Display/Registered Contourer to verify that the registrations are consistent (also verify image orientation and patient orientation labels)

Registered Contourer Actor

Preparation

  1. Import images from patient 2,3,4,5,6,10
  2. For each patient, use registration from registrator, contour on both primary and secondary images.
  3. Contours must be saved in the FoR of the Primary (CT Image).
  4. Export contours to the Archive.
  5. Please indicate the vendor of the Registered Contourer Actor in the Series Description of the RT Structure Set stored to the Archive.

Testing

For each Peer (Registered Contourer, Registered Display):

  1. Tester imports images, registrations and contours from Peer's corresponding patient
  2. Tester shows it can read in and apply the registration
  3. Tester shows it can display contours on both images (using registration above)
  4. Tester draws contours on one of the images
  5. Tester moves contours on to other images (using imported registration)
  6. Tester exports these contours (now in other FoR than drawn) to PACS
  7. Peer imports images and contours and displays them correctly (name, colour, shape, position)

Registered Display Actor

Preparation

  1. Tester imports images, structure set, and SRO for patient 2,3,4,5,6,10

Testing For each peer do the following

  1. Tester shows it can read in and apply the registration
  2. Tester shows it can display contours on both image Series using imported SRO
  3. Side-by-side comparison of Registered Display and Registered Contourer is used to verify interoperable transfer.

Registered Dose Display Actor

Note that the creation of dose objects may not necessarily be done by a peer being tested for this profile, e.g. BRTO Dosimetric Planner.

Preparation (Not actually a requirement for the vendor being tested to pass the profile, but encouraged if possible to create the test data needed for testing)

  1. Use patient 2,3,4 or 10
  2. Create registration between primary and secondary image
  3. Create contours, plan & dose on primary images
  4. Export registration, contours & dose to Archive

Testing This is exactly the same as for Registered Display, but tester should also demonstrate the ability to correctly display dose on registered (i.e., secondary images) using SRO



Supported Transactions

Actors Transactions Optionality Section
Archive Modality Images Stored R RAD 4.8
Creator Images Stored R RAD 4.18
Registered Structure Set Storage R MMRO-3
Spatial Registration-II Storage R MMRO-II-1
Spatial Registration-II Retrieval R MMRO-II-2
Registered Structure Set Retrieval R MMRO-4
Registered Dose Retrieve R MMRO-5
Registrator Modality Images Stored R RAD 4.8
Creator Images Stored O RAD 4.18
Spatial Registration-II Storage O MMRO-II-1
Spatial Registration-II Retrieval R MMRO-II-2
Registered Contourer Modality Images Stored R RAD 4.8
Registered Structure Set Storage R MMRO-3
Registered Structure Set Retrieval R MMRO-4
Spatial Registration-II Retrieval R MMRO-II-2
Registered Display Modality Images Stored R RAD 4.8
Registered Structure Set Retrieval R MMRO-4
Spatial Registration-II Retrieval R MMRO-II-2
Registered Dose Display Modality Images Stored R RAD 4.8
Registered Structure Set Retrieval R MMRO-4
Registered Dose Retrieval R MMRO-5
Spatial Registration-II Retrieval R MMRO-II-2